User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: B&K 2131: Digital Frequency Analyzer DATE: 24-Sep-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 8 NUMBER OF LINES: 92 CONFIGURATION: Fluke 5700A STANDARD: B&K 4220 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P M F W 1.002 ASK+ X 1.003 HEAD INITIAL CONDITIONS 1.004 DISP Connect the UUT as follows: 1.004 DISP 5700 A V/OHM/A OUTPUTS ............... UUT DIRECT IN 1.005 CALL 2131 Attenuator Sub 1.006 CALL 2131 Linearity Sub 1.007 CALL 2131 Frequency Response Sub 1.008 HEAD {* "A" Weighting Filter *} 1.009 JMP 2.001 1.010 EVAL 2.001 DISP Set the UUT as follows: 2.001 DISP FREQUENCY RANGE .................... 1.6 - 1.25k 2.001 DISP CENTER FREQUENCY ................... 1.00 KHZ 2.001 DISP A WEIGHTING ........................ On 2.002 DISP Set UUT attenuator for 130dB full scale. 2.003 5700 1V 1kH S 2W 2.004 MEMI Enter the UUT LEVEL indication in dB. 2.005 MEME >1 2.006 DISP Set the UUT as follows: 2.006 DISP CENTER FREQUENCY ................... 10 HZ 2.006 DISP INPUT ATTEN. ....................... 100 dB 2.007 5700 1V 10H S 2W 2.008 MEMI Enter the UUT LEVEL indication in dB. 2.009 MATH mem1= m[1] - 70 2.010 MEME 2.011 MEMC d 5.0U 10H #! Test Tol 5, Sys Tol 0.00065, TUR 7692.308 (>= 4.00). 3.001 DISP Set the UUT as follows: 3.001 DISP CENTER FREQUENCY ................... 100 HZ 3.001 DISP INPUT ATTEN. ....................... 130 dB 3.002 5700 1V 100H S 2W 3.003 MEMI Enter the UUT LEVEL indication in dB. 3.004 MATH mem1= m[1] - 18 3.005 MEME 3.006 MEMC d 5.0U 100H #! Test Tol 5, Sys Tol 8.2e-005, TUR 60975.610 (>= 4.00). 4.001 DISP Set the UUT as follows: 4.001 DISP FREQUENCY RANGE .................... 25 - 20k 4.001 DISP CENTER FREQUENCY ................... 5.00 KHZ 4.002 5700 1V 5kH S 2W 4.003 MEMI Enter the UUT LEVEL indication in dB. 4.004 MEME 4.005 MEME <1 4.006 MEMC d 5.0U 5kH #! Test Tol 5, Sys Tol 8.2e-005, TUR 60975.610 (>= 4.00). 5.001 DISP Set the UUT as follows: 5.001 DISP CENTER FREQUENCY ................... 20 KHZ 5.002 5700 1V 20kH S 2W 5.003 MEMI Enter the UUT LEVEL indication in dB. 5.004 MATH mem1= m[1] - 12.5 5.005 MEME 5.006 MEMC d 5.0U 20kH #! Test Tol 5, Sys Tol 8.2e-005, TUR 60975.610 (>= 4.00). 6.001 HEAD {* Preamp Input *} 6.002 JMP 7.001 6.003 EVAL 7.001 STD B&K 4220 7.002 DISP Disconnect the 5700A from the UUT DIRECT INPUT> 7.003 DISP Connect microphone capsule (DTB# 13-53-2) to the 7.003 DISP UUT PREAMP INPUT. 7.004 DISP Set the UUT as follows: 7.004 DISP INPUT .............................. Preamp 7.004 DISP FREQUNCY RANGE ..................... 1.6 - 1.25k 7.004 DISP CENTER FREQUENCY ................... 250 HZ 7.005 DISP Insert the UUT microphone into the 7.005 DISP Model 4220 Piston Phone. 7.006 DISP Turn on the Piston Phone. 7.007 MEMI Enter the UUT LEVEL indication. 7.008 MATH mem1 = 123.9 7.009 MEME 7.010 MEMC d 0.2U #! T.U.R. not calculated because System Uncertainty not available. 8.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 1 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.